Measuring Equipment Sound / Noise Kapseln MK301

MK301_klMK301 

1/4" Condenser
Measuring Microphone Capsule 
Free-field
Frequency range 5 Hz - 100 kHz,
Dynamic range 35 dB(A) - 168 dB

 

The 1/4” MK 301 measuring microphone capsule is designed for acoustical measurements in research and development and also for industrial use. Some applications include audiometry, the measurement of building acoustics and noise levels.
  • Frequency range 5 Hz to 100 kHz, free-field
  • Sound pressure levels up to 168 dB

The microphone capsule is designed and very carefully constructed to ensure excellent long-time stability of the electroacoustical parameters. All essential components including the diaphragm are made of nickel and the diaphragm is bonded by a special galvanic process. The rear electrode is insulated from the capsule housing by a quartz glass plate. The diaphragm is protected against mechanical damage by a protection grid.

The equalization of the static air pressure between the inside and outside of the capsule is by means of a capillary tube side-vented.
The individual pressure frequency response curves can be convenient measured by using a calibration grid, e.g. UA 0033 by means of the adaptor DB 0264.
The capsule uses the international standard thread 60 UNS which ensures compatibility with a wide range of calibration equipment and measurement devices available from many manufacturers.
Measuring microphone cartridge MK 301 and measurement microphone preamplifier MV 302/ MV 301 can be calibrated.

Maintenance and Upkeep

In order to maintain its functioning the measurement microphone capsule should be protected against mechanical damage. It should be disconnected from its power source and completely checked for any pollution in regular intervals that have to be defined depending on the operating conditions.
After removal of the protection grid the pollution within the grid and on the diaphragm should be cleaned very carefully with a soft clean cloth or brush.
The measurement microphone capsule is not suitable for use in chemical aggressive conditions and in conductible dust. Condensation must be avoided.

Messmikrofonkapsel
MK 301
21.31- 96.124 PTB-Zulassung-Nr.zur amtlichen Eichung
Wandlertyp
Kapazitiver Druckempfänger
*
Frequenzbereich des Freifeldübertragungsmaßes
ohne Schutzkappe
5 Hz ... 100 kHz, (± 2 dB)
*
Feld-Leerlauf-Übertragungsfaktor
5 mV/Pa
Grenzschalldruckpegel für 3 % Klirrfaktor bei 1 kHz
168 dB
Eigenrauschen mit Vorverstärker MV 302
35 dBA
Polarisationsspannung
200 V
*
Kapazität mit Polarisationsspannung bei 1 kHz
6,8 pF
Arbeitstemperaturbereich
-50 ... +110 °C
Temperaturkoeffizient
≤ 0,01 dB/K
Statischer Druckkoeffizient
-1x10-5 dB/Pa
Durchmesser          mit Schutzkappe
7,0 ± 0,02 mm
Durchmesser      ohne Schutzkappe
6,35 ± 0,02 mm
Höhe
9,7 mm
Gewicht
2 g
Gewinde für Vorverstärker
5,7 mm 60 UNS
Gewinde für Schutzkappe
6,35 mm 60 UNS
* individuell kalibriert

 

Frequenzgänge

freq

Polardiagramme

pol_01_k pol_02_k pol_03_k



Messmikrofonkapsel MK 301 im Holzetui
Best.Nr. 311109